IN-VITRO COMPARISON OF 2 BALLISTIC SYSTEMS FOR ENDOSCOPIC STONE DISINTEGRATION

Citation
G. Haupt et al., IN-VITRO COMPARISON OF 2 BALLISTIC SYSTEMS FOR ENDOSCOPIC STONE DISINTEGRATION, Journal of endourology, 10(5), 1996, pp. 417-420
Citations number
43
Categorie Soggetti
Urology & Nephrology
Journal title
ISSN journal
08927790
Volume
10
Issue
5
Year of publication
1996
Pages
417 - 420
Database
ISI
SICI code
0892-7790(1996)10:5<417:ICO2BS>2.0.ZU;2-B
Abstract
Endoscopic stone disintegration can be performed with various lithotri psy techniques. We compared two ballistic systems in vitro: the Swiss Lithoclast and the Combilith (Olympus), Measurements of the probe move ment; i.e., maximum velocity and displacement, were performed using an optical high-speed technique based on a frame transfer camera. The Li thoclast and Combilith both achieve excellent probe velocities. Howeve r, the Lithoclast achieves equal velocity with less probe displacement .