Hd. Hummler et al., PATIENT-TRIGGERED VENTILATION IN NEONATES - COMPARISON OF A FLOW-TRIGGERED AND AN IMPEDANCE-TRIGGERED SYSTEM, American journal of respiratory and critical care medicine, 154(4), 1996, pp. 1049-1054
Citations number
22
Categorie Soggetti
Emergency Medicine & Critical Care","Respiratory System
We conducted a study with the objective of comparing the performance o
f two different systems for patient-triggered ventilation in neonates
(impedance versus flow/volume-triggered) by measuring response time, a
utotrigger and trigger failure rates, ventilation, and gas exchange. T
he two ventilator systems were applied in random order in 10 preterm n
eonates (median gestational age: 30.5 wk; range: 27 to 34 wk; body wei
ght: 1,266 g; range: 840 to 2,240 g) using identical ventilator settin
gs. The median (range) response time was 169 (98 to 305) ms for the im
pedance system and 115 (79 to 184) ms for the flow/volume system (p <
0.01). The longer and more variable response time of the impedance sys
tem was secondary to a phase lag of the impedance signal caused by che
st wall distortion. Although 13.7 (0.2 to 29.4)% of mechanical breaths
were autotriggered with the impedance system, there were no autotrigg
ered breaths using the flow/volume system (p < 0.01). The rate of trig
ger failures was not significantly different with the two systems, at
1.2 (0 to 4.4)% (impedance) versus 3.1 (0 to 6.4)% (flow/volume). Minu
te ventilation was smaller with the impedance system (p < 0.001), beca
use of the larger number of breaths triggered late in inspiration or d
uring expiration. We conclude that the flow/volume-triggered system is
less prone to autotriggering and has a shorter and more consistent re
sponse time than the impedance-triggered system. The impedance-trigger
ed system is more susceptible to artifacts and chest wail distortion.