A flow injection (FI) analytical method for the simultaneous assay of
nitrite, nitrate and chloride in meat products is reported, The method
is based on the potentiometric determination of chloride using a tubu
lar ISE and on the spectrophotometric determination of nitrite. The FI
system consisted in splitting the flow after potentiometric detection
using a tubular detector and the subsequent confluence of the flow be
fore reaching the spectrophotometric detector, This allowed the reduct
ion of nitrate to nitrite in part of the sample plug on an on-line cop
per cadmium reductor column. Since each channel had a different reside
nce time, two peaks were obtained for nitrite and nitrite plus nitrate
, Spectrophotometric determination was made after a diazotization-coup
ling reaction. The results obtained were in good agreement with refere
nce procedures and showed adequate precision (RSDs less than 6% for ch
loride and nitrite and 2% for nitrate). A high sampling rate was obtai
ned (120 determinations per hour corresponding to 40 samples per hour)
.