ANALYSIS OF THE OUTPUT SIGNAL WAVE-FORM AND PERFORMANCES OF SEMIINSULATING GAAS PARTICLE DETECTORS

Citation
Aa. Quaranta et al., ANALYSIS OF THE OUTPUT SIGNAL WAVE-FORM AND PERFORMANCES OF SEMIINSULATING GAAS PARTICLE DETECTORS, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 380(1-2), 1996, pp. 201-204
Citations number
22
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
380
Issue
1-2
Year of publication
1996
Pages
201 - 204
Database
ISI
SICI code
0168-9002(1996)380:1-2<201:AOTOSW>2.0.ZU;2-D
Abstract
This paper presents an extensive analysis of the signals at the output of a charge sensitive amplifier delivered by a semi-insulating liquid -encapsulated Czochralski grown GaAs particle detector, irradiated wit h 5.48 MeV alpha particles. When the detector is operated below full d epletion the shape of the output signal can be explained by the equiva lent circuit of the detector itself, while above full depletion it is affected only by trapping-detrapping processes and it is well interpre ted in terms of trapping and detrapping times of deep level traps pres ent in the materials. Spectral lines of alpha particles from an Am-241 source have shown a resolution of 1.1% at room temperature.