LOW-NOISE JFETS FOR ROOM-TEMPERATURE X-RAY-DETECTORS

Citation
Mw. Lund et al., LOW-NOISE JFETS FOR ROOM-TEMPERATURE X-RAY-DETECTORS, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 380(1-2), 1996, pp. 318-322
Citations number
21
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
380
Issue
1-2
Year of publication
1996
Pages
318 - 322
Database
ISI
SICI code
0168-9002(1996)380:1-2<318:LJFRX>2.0.ZU;2-0
Abstract
We have developed a series of JFETs specifically designed for energy d ispersive spectrometer preamplifiers. These have very low channel and generation-recombination noise both at room temperature and at low tem perature. We discuss the design of JFETs for low noise, present some r esults using these with room temperature operation, and discuss furthe r possible improvements.