Se. Vandenbosch et al., EXPERIENCES WITH PACKED CAPILLARY ELECTROCHROMATOGRAPHY AT AMBIENT-PRESSURE, Journal of chromatography, 755(2), 1996, pp. 165-177
Citations number
21
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
The practical execution of packed capillary electrochromatography is e
valuated. The most serious problem is the generation of air bubbles in
the system. This occurs especially in frits which were fabricated by
glueing particles differing from the packing particles. No air bubbles
were generated with frits which were prepared by locally sintering a
narrow band of the reversed-phase packing by means of a heating filame
nt, while flushing the column with water. In this way electrochromatog
raphy can be performed simply without pressurizing the buffer vials, p
roviding the mobile phase is frequently degassed with helium. The perf
ormance under electrically driven conditions appears to be excellent a
nd for 5 mu m reversed-phase particles reduced plate heights down to 1
.8 were achieved. Compared to pressure driven conditions, 25-40% small
er plate heights were obtained under electrically driven conditions.