SIGNAL-PROCESSING FOR CDTE DETECTORS USING A FAST SIGNAL DIGITIZING TECHNIQUE

Citation
H. Takahashi et al., SIGNAL-PROCESSING FOR CDTE DETECTORS USING A FAST SIGNAL DIGITIZING TECHNIQUE, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 380(1-2), 1996, pp. 381-384
Citations number
9
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
380
Issue
1-2
Year of publication
1996
Pages
381 - 384
Database
ISI
SICI code
0168-9002(1996)380:1-2<381:SFCDUA>2.0.ZU;2-0
Abstract
To suppress the charge collection loss variation in CdTe detectors, a signal processing method has been developed using a fast signal digiti zing technique. Signal pulse waveforms were described as a modeling fu nction. Then a set of model waveform patterns has been calculated for reference patterns. The most probable waveform pattern was selected by the pattern matching method which assures a finite calculation time. Obtained residue values were used for the further selection in the cor rection method. An energy resolution of 8.9 keV was obtained after the correction and selection of measured pulse signals for the Cs-137 662 keV photopeak. This method can be applied to a detector having any el ectrode configuration if the signal pulses from that configuration are given.