H. Takahashi et al., SIGNAL-PROCESSING FOR CDTE DETECTORS USING A FAST SIGNAL DIGITIZING TECHNIQUE, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 380(1-2), 1996, pp. 381-384
To suppress the charge collection loss variation in CdTe detectors, a
signal processing method has been developed using a fast signal digiti
zing technique. Signal pulse waveforms were described as a modeling fu
nction. Then a set of model waveform patterns has been calculated for
reference patterns. The most probable waveform pattern was selected by
the pattern matching method which assures a finite calculation time.
Obtained residue values were used for the further selection in the cor
rection method. An energy resolution of 8.9 keV was obtained after the
correction and selection of measured pulse signals for the Cs-137 662
keV photopeak. This method can be applied to a detector having any el
ectrode configuration if the signal pulses from that configuration are
given.