P. Aggarwal et D. Dollimore, THE USE OF THE TIME-TEMPERATURE SIGNAL IN THE THERMOGRAVIMETRY UNIT ILLUSTRATED BY ANALYSIS OF INDIUM, ALUMINUM AND ZINC, Instrumentation science & technology, 24(4), 1996, pp. 299-305
The use of a time-temperature signal present in a simultaneous TG-DTA
analyser is described. The melting behaviours, on the DTA signal, of i
ndium, aluminium and zinc are reported as the basis to illustrate the
performance of the equipment for acquisition of relevant data. This on
ly requires the signal from the sample thermocouple to be recorded. By
noting the perturbation of the time - temperature signal from the pro
jected baseline, a curve similar to the conventional DTA curve can be
obtained.