Several earlier studies have indicated that bileaflet mechanical heart
valves behave irregularly at low cardiac output and low pulse rate co
nditions, and that their hydrodynamic performances are generally inade
quate. The authors conducted in vitro experiments in a pulsatile mock
circulatory loop to compare the performance of the St. Jude Medical (S
JM) valve and a long body bileaflet prosthesis recently introduced by
Medical Carbon Research Institute (MCRI) (Austin, TX). The new MCRI me
chanical heart valve model was designed with emphasis on improved hydr
odynamic efficacy by introducing a long body with parallel leaflets an
d by leaflets increasing the flow area. Experimental studies were cond
ucted on five test valves (MCRI 19 mm, MCRI 25 mm, SJM 19 mm, SJM 23 m
m, and SJM 29 mm) with cardiac outputs of 2.0, 2.5, 3.0, and 3.5 L/min
at a pulse rate of 40 beats/min, and 3.5, 4.0, 4.5, and 5.0 L/min at
a pulse rate of 70 beats/min. Transvalvular pressure drop and closure
volume were assessed by measuring the instantaneous ventricular and ao
rtic pressures and aortic flow. The leaflet motions of the tested valv
es were observed by direct video recording using a charge coupled devi
ce camera while the flow measurements were being conducted. Testing un
der simulated physiologic ventricular and aortic pressure waveforms, t
he results of this study show that the MCRI bileaflets remained fully
open during the entire ejection phase, even at very low cardiac output
conditions (2.0 L/min). The closure volume (defined as percentage of
forward flow volume) increased with decreasing cardiac output, as repo
rted earlier by others. Comparative results also indicate that the MCR
I design has nearly a two size pressure drop advantage over the SJM, w
ith significantly smaller closure volume.