GROWTH-HORMONE ASSAYS - EARLY TO LATEST TEST GENERATIONS COMPARED

Citation
M. Lhermitebaleriaux et al., GROWTH-HORMONE ASSAYS - EARLY TO LATEST TEST GENERATIONS COMPARED, Clinical chemistry, 42(11), 1996, pp. 1789-1795
Citations number
14
Categorie Soggetti
Chemistry Medicinal
Journal title
ISSN journal
00099147
Volume
42
Issue
11
Year of publication
1996
Pages
1789 - 1795
Database
ISI
SICI code
0009-9147(1996)42:11<1789:GA-ETL>2.0.ZU;2-R
Abstract
We compared the data from four growth hormone (GH) immunoassays for an alyzing 24-h GH profiles in four apparently normal subjects and four o bese subjects (508 serum samples), The detection limit was 0.02 mu g/L for one immunochemiluminometric assay (ICMA), 0.1 mu g/L for two IRMA s, and 0.4 mu g/L for one RIA. All GH pulses with a peak ICMA value >1 mu g/L were detected by each of the other methods. Overall, the corre lation coefficient between the values obtained with all four assays ex ceeded 0.90. However, for GH concentrations less than or equal to 0.25 mu g/L, acceptable concordance (r(2) greater than or equal to 0.80) w as reached only between the ICMA and one IRMA; between the ICMA. and t he RIA, concordance was acceptable only for GH concentrations greater than or equal to 10 mu g/L. In the normal subjects, the percentage of undetectable values was 0% with the ICMA but 29% with one of the IRMAs ; in obese subjects, the corresponding values were 12% and 38%.