RETROFILTRATION RATES IN HIGH-FLUX HOLLOW-FIBER HEMODIALYZERS - ANALYSIS OF CLINICAL-DATA

Citation
A. Wupper et al., RETROFILTRATION RATES IN HIGH-FLUX HOLLOW-FIBER HEMODIALYZERS - ANALYSIS OF CLINICAL-DATA, Journal of membrane science, 121(1), 1996, pp. 109-116
Citations number
18
Categorie Soggetti
Engineering, Chemical","Polymer Sciences
Journal title
ISSN journal
03767388
Volume
121
Issue
1
Year of publication
1996
Pages
109 - 116
Database
ISI
SICI code
0376-7388(1996)121:1<109:RRIHHH>2.0.ZU;2-D
Abstract
Clinical data obtained with high-flux hollow fiber hemodialyzers opera ting under the condition of zero net volume flow of filtrate are analy zed theoretically to obtain information about the profile of the radia l volume flow density, the concentration of an impermeable solute alon g the axis of the hollow fibers, and the extent of retrofiltration. Th is information is of practical interest. The clinical data refer to di alyzers differing only in the diameter of the hollow fibers.