POTENTIOMETRIC DETECTION OF CARBOXYLIC-ACIDS BY FLOW-INJECTION ANALYSIS USING A TUNGSTEN-OXIDE ELECTRODE

Citation
Zl. Chen et Pw. Alexander, POTENTIOMETRIC DETECTION OF CARBOXYLIC-ACIDS BY FLOW-INJECTION ANALYSIS USING A TUNGSTEN-OXIDE ELECTRODE, Analytica chimica acta, 332(2-3), 1996, pp. 187-192
Citations number
16
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032670
Volume
332
Issue
2-3
Year of publication
1996
Pages
187 - 192
Database
ISI
SICI code
0003-2670(1996)332:2-3<187:PDOCBF>2.0.ZU;2-N
Abstract
Potentiometric flow injection (PFI) analysis of carboxylic acids based on a tungsten oxide electrode is described. The electrode is shown to offer high sensitivity and a stable baseline for the detection of car boxylic acids. The parameters affecting the electrode response such as the concentration of buffer, the pH and flow rate were studied in det ail. The electrode exhibited a linear response for propionic acid in t he concentration range of 2.5x10(-4)M-5x10(-5)M using a 5x10(-4)M phos phate buffer at pH 6.5 as a carrier. The detection limits were found t o be in the range of 1.0x10(-5)M-5x10(-5)M with a 20 mu l injection.