CONTRALATERAL LOOP SNARE REMOVAL OF A RUPTURED AND ENTRAPPED ANGIOPLASTY BALLOON

Citation
Ma. Braun et al., CONTRALATERAL LOOP SNARE REMOVAL OF A RUPTURED AND ENTRAPPED ANGIOPLASTY BALLOON, Cardiovascular and interventional radiology, 19(6), 1996, pp. 428-430
Citations number
11
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
01741551
Volume
19
Issue
6
Year of publication
1996
Pages
428 - 430
Database
ISI
SICI code
0174-1551(1996)19:6<428:CLSROA>2.0.ZU;2-N
Abstract
We describe a technique that extracted a ruptured angioplasty balloon which had become entrapped by a calcified left common iliac artery ste nosis. The balloon catheter had been advanced crossover from the right and could not be retracted directly into a sheath across the aortic b ifurcation. Therefore, a guidewire was inserted through the balloon ca theter and captured by a loop snare advanced from the left femoral art ery, The loop snare was also used to free the balloon wings from the s tenosis. The balloon was then pulled into a 10 Fr sheath and removed a s a unit with the sheath.