FLOW-INJECTION STOPPED-FLOW SPECTROFLUOROMETRIC KINETIC DETERMINATIONOF PARACETAMOL BASED ON ITS OXIDATION REACTION BY HEXACYANOFERRATE(III)

Citation
Jam. Pulgarin et Lfg. Bermejo, FLOW-INJECTION STOPPED-FLOW SPECTROFLUOROMETRIC KINETIC DETERMINATIONOF PARACETAMOL BASED ON ITS OXIDATION REACTION BY HEXACYANOFERRATE(III), Analytica chimica acta, 333(1-2), 1996, pp. 59-69
Citations number
35
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032670
Volume
333
Issue
1-2
Year of publication
1996
Pages
59 - 69
Database
ISI
SICI code
0003-2670(1996)333:1-2<59:FSSKD>2.0.ZU;2-N
Abstract
A simple and economic system, easily adaptable to any spectrophotomete r or spectrofluorimeter in order to carry out kinetic measurements by means of the FIA stopped-flow technique, is described. The system cons ists of a 12 bit analogical digital converter that receives analogue s ignals at -10 and +10V and enables the electrical supply to the perist altic pump to be interrupted. A stopped-flow method with fluorescence detection for the determination of paracetamol (N-acetyl-p-aminophenol ), based on its oxidation with hexacyanoferrate(III), is described. A kinetic study of the reaction is developed measuring the initial rate of change of the fluorescence intensity of the oxidized product formed at 241 and 426 nm excitation and emission wavelengths, respectively. The calibration graph is linear in the range of concentrations 0.5-15. 0 mu g/ml. The relative standard deviation is 2.5%. The method was app lied to authentic pharmaceutical preparations and was found to be sati sfactory. The effect of same possible interferences has been studied.