Mj. Mooney et al., MONITORING AND CORRECTION OF MULTIPLE SCATTER DURING CLINICAL COMPTONSCATTER DENSITOMETRY MEASUREMENTS, Physics in medicine and biology, 41(11), 1996, pp. 2399-2410
A system for monitoring multiple scatter during a clinical Compton sca
tter densitometry measurement of bone density is described. Multiple s
catter from the measurement site was measured using a supplementary co
llimated detector positioned so that only multiply scattered photons C
ould enter the detector. The data from the detector were used to form
a multiple-scatter correction factor (mcf) to correct for the bias att
ributed to multiple scatter. The results of experimental and computer
simulations are presented which demonstrate the relationship between t
he multiple-scatter reading and calculated mcf values. The influence o
f bone size on the values of mcf in large measurement sites, such as t
he femoral neck, was shown to he negligible. A simulation was used to
produce a visualization of the multiple scatter in order to ascertain
the optimum position of the supplementary detector. This technique was
shown to be a rapid and accurate method of measuring the multiple-sca
tter bias and suitable for use during clinical CSD measurements.