Measurements of atmospheric ozone at concentrations typical of the fre
e troposphere have been compared for ECC sondes and a UV-absorption ph
otometer, using a Bendix chemiluminescent analyser as a transfer stand
ard. Comparisons were conducted in the laboratory and in the troposphe
ric part of the atmosphere. It was found that the measurements agreed
to within 4% provided that the background current for the ECC sonde wa
s measured before exposure to ozone in the preparation procedure, and
was assumed to be constant throughout the sonde flight. These results
confirm those of earlier experiments and mean that the methods current
ly used to correct for the background current in the troposphere need
to be revised.