P. Bilski et al., PROPERTIES OF DIFFERENT THIN-LAYER LIF-MG,CU,P TL DETECTORS FOR BETA-DOSIMETRY, Radiation protection dosimetry, 66(1-4), 1996, pp. 101-104
Citations number
6
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
Dosimetric properties of two different types of thin TL detectors, des
igned to measure personal doses at a depth of 7 mg.cm(-2) in tissue, h
ave been compared. At Riso National Laboratory detectors were prepared
by fixing LiF:Mg,Cu,P GR-200 thin detectors to 0.7 mm thick Al pieces
. At the Institute of Nuclear Physics, a thin layer of LiF:Mg,Cup phos
phor of effective thickness 8.5 mg.cm(-2) was sintered to a base of un
doped LiF. Beta energy and angular response, detection threshold, stab
ility, etc. were investigated for the two detector types. Results show
that both detectors possess good energy characteristic and that their
sensitivity is sufficient to measure doses below 100 mu Gy reliably.
Hence, they are suitable for practical applications in extremity dosim
etry.