ORGAN DOSE DETERMINATIONS OF X-RAY EXAMINATIONS USING TL DETECTORS FOR VERIFICATION OF COMPUTED DOSES

Citation
M. Toivonen et al., ORGAN DOSE DETERMINATIONS OF X-RAY EXAMINATIONS USING TL DETECTORS FOR VERIFICATION OF COMPUTED DOSES, Radiation protection dosimetry, 66(1-4), 1996, pp. 289-294
Citations number
15
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
66
Issue
1-4
Year of publication
1996
Part
2
Pages
289 - 294
Database
ISI
SICI code
0144-8420(1996)66:1-4<289:ODDOXE>2.0.ZU;2-6
Abstract
The organ and effective doses computed with a new flexible program wer e compared in four X ray examinations with the doses determined with L F TL detectors in an Alderson-Rando phantom. The correlation between t he two methods was rather good (within 25%) for most results of the co mparisons (75% of 45 results), but rather poor (discrepancy up to 140% ) for the low and high density organs (lung, bone). An upward trend wa s also obtained for the doses to colon and some other organs. The reas ons for the discrepancies were thought to be due to the approximate de nsity and attenuation corrections of the computer program for bone and lung tissues.