F. Pernicka et Hq. Linh, USE OF UNIVERSAL FUNCTIONAL OPTIMIZATION FOR TL GLOW CURVE ANALYSIS, Radiation protection dosimetry, 66(1-4), 1996, pp. 437-440
Citations number
4
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
The effective use of any TL instrument requires an efficient software
package to be able to fulfil different tasks required by research and
practical applications. One of the standard features of the package us
ed at the NPI Prague is the application of the interactive modular sys
tem Universal Functional Optimisation (UFO) for glow curve deconvoluti
on. The whole system has been tested on standard glow curves using dif
ferent models of the TL process (a single peak described by the Podgor
sak approximation, first order kinetics and/or general order kinetics)
. Calculated values of basic TL parameters (E and s) show a good agree
ment with the results obtained by other authors. The main advantage of
the system is in its modularity that enables flexible changes in the
TL model and mathematical procedures of the glow curve analysis.