QUANTITATIVE ELECTRON-PARAMAGNETIC-RESONANCE (EPR) SPECTROMETRY WITH A TE(104) DOUBLE RECTANGULAR CAVITY .1. A SIMPLE ALIGNMENT PROCEDURE FOR THE PRECISION POSITIONING OF THE SAMPLE

Citation
M. Mazur et al., QUANTITATIVE ELECTRON-PARAMAGNETIC-RESONANCE (EPR) SPECTROMETRY WITH A TE(104) DOUBLE RECTANGULAR CAVITY .1. A SIMPLE ALIGNMENT PROCEDURE FOR THE PRECISION POSITIONING OF THE SAMPLE, Analytica chimica acta, 333(3), 1996, pp. 249-252
Citations number
18
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032670
Volume
333
Issue
3
Year of publication
1996
Pages
249 - 252
Database
ISI
SICI code
0003-2670(1996)333:3<249:QE(SWA>2.0.ZU;2-Y
Abstract
The Bruker TE(104) double rectangular cavity is commonly utilized for quantitative electron paramagnetic resonance (EPR) spectrometric measu rements. The sample, together with an appropriate standard, can be stu died simultaneously within such a cavity under identical states of mac hine tuning and parameter settings. However, even under such condition s, there still remain sources of error arising from sample size, shape and positioning of the sample tube within the cavity, In this paper s uch sources of error have been investigated and a procedure to minimiz e the errors by precision insertion of the sample tubes within the TE( 104) cavity is described.