The usefulness of the Fluorotron Master(R) fluorophotometer fitted wit
h a small animal adapter was evaluated with tree shrews, We also deter
mined that the optimal measurement parameters for these animals, which
are regarded as promising for experimental use in ophthalmic research
, are: fluorescein-Na concentration, 0.5x10(-9) to 1.0x10(-6) g/mL; fl
uorescein-Na dosage, 2 mg/kg; measurement time, 30 minutes after injec
tion of fluorescein-Na into the ocular compartments. Results indicated
that ultrafiltration for the measurement of protein-unbound fluoresce
in can be done with a hematocrit tube and a minimal blood sample, ther
eby reducing the impact of the sampling on the animal's general condit
ion, The Fluorotron Master(R) with the small animal adapter can offer
advantages in estimating the blood-ocular barrier permeability using d
isease models in tree shrews.