M. Elsner et al., INTRACORONARY LOSS OF BALLOON-MOUNTED STENTS - SUCCESSFUL RETRIEVAL WITH A 2 MM MICROSNARE-DEVICE, Catheterization and cardiovascular diagnosis, 39(3), 1996, pp. 271-276
Intracoronary loss of balloon-mounted stents prior to deployment is a
rare event but may result in clinically relevant cardiac ischemia or p
eripheral embolization during rescue attempts. In a consecutive series
of nearly 400 patients undergoing elective stenting, we encountered s
ix instances of intracoronary misplacement. We describe four cases of
successful percutaneous retrieval using a 2 mm closed loop nitinol dev
ice (''Micro-snare''). Our proposed method of extraction seems to hold
advantages over those previously described. (C) 1996 Wiley-Liss Inc.