Eyd. Chow et Kk. Leung, EVALUATING THE CELL-DYN(R)-3500 HEMATOLOGY ANALYZER IN AN ACUTE GENERAL-HOSPITAL, Clinical and laboratory haematology, 18(3), 1996, pp. 187-193
The CELL-DYN(R) 3500 (CD3500) (Abbott Diagnostics Division, Santa Clar
a, CA, USA) is a multiparameter, automated haematology analyser system
capable of producing 22 haematological parameters including a screeni
ng five-part differential. The evaluation was performed in a recently
expanded acute care general hospital. Studies of linearity, carryover
and precision were acceptable and within manufacturer's stated limits,
Correlation of CD3500 with manual differential counts were good excep
t for the basophil count, The white cell nagging system had a sensitiv
ity of 83.6% and specificity of 88.9% with 85.6% agreement. The platel
et nagging system also had comparable sensitivity and specificity rate
of 88.4% and 94.5% and agreement rate of 92.2%. The overall false pos
itive rate and false negative rate for both the white cell and platele
t flags combined were 8.6% and 10.3%. All samples provided a different
ial with no rejection experienced. The MAPSS technology, together with
the extended lyse mode and the investigational nine-part differential
provides potential for future development in WBC differential.