EVALUATING THE CELL-DYN(R)-3500 HEMATOLOGY ANALYZER IN AN ACUTE GENERAL-HOSPITAL

Authors
Citation
Eyd. Chow et Kk. Leung, EVALUATING THE CELL-DYN(R)-3500 HEMATOLOGY ANALYZER IN AN ACUTE GENERAL-HOSPITAL, Clinical and laboratory haematology, 18(3), 1996, pp. 187-193
Citations number
11
Categorie Soggetti
Hematology
ISSN journal
01419854
Volume
18
Issue
3
Year of publication
1996
Pages
187 - 193
Database
ISI
SICI code
0141-9854(1996)18:3<187:ETCHAI>2.0.ZU;2-5
Abstract
The CELL-DYN(R) 3500 (CD3500) (Abbott Diagnostics Division, Santa Clar a, CA, USA) is a multiparameter, automated haematology analyser system capable of producing 22 haematological parameters including a screeni ng five-part differential. The evaluation was performed in a recently expanded acute care general hospital. Studies of linearity, carryover and precision were acceptable and within manufacturer's stated limits, Correlation of CD3500 with manual differential counts were good excep t for the basophil count, The white cell nagging system had a sensitiv ity of 83.6% and specificity of 88.9% with 85.6% agreement. The platel et nagging system also had comparable sensitivity and specificity rate of 88.4% and 94.5% and agreement rate of 92.2%. The overall false pos itive rate and false negative rate for both the white cell and platele t flags combined were 8.6% and 10.3%. All samples provided a different ial with no rejection experienced. The MAPSS technology, together with the extended lyse mode and the investigational nine-part differential provides potential for future development in WBC differential.