AUTOMATIC CHROMIUM (III) (VI) PRECONCENTRATION AND SPECIATION BY FLOWINJECTION-FLAME AAS/

Citation
Ms. Jimenez et al., AUTOMATIC CHROMIUM (III) (VI) PRECONCENTRATION AND SPECIATION BY FLOWINJECTION-FLAME AAS/, Atomic spectroscopy, 17(5), 1996, pp. 201-207
Citations number
38
Categorie Soggetti
Spectroscopy
Journal title
ISSN journal
01955373
Volume
17
Issue
5
Year of publication
1996
Pages
201 - 207
Database
ISI
SICI code
0195-5373(1996)17:5<201:AC((PA>2.0.ZU;2-B
Abstract
A sensitive, rapid flow injection flame AAS method for the simultaneou s determination of Cr(III) and Cr(VI) following ion-exchange preconcen tration is proposed. Chromium(III) is separated and preconcentrated on a cation-exchange resin, Amberlite IR-120, and Cr(VI) on Amberlite IR A-400. Preconcentration parameters (column size, eluent, loading and e lution flow rate) were optimized, and the phase transfer, retention an d enrichment factor for both species determined. The linear dynamic ra nges achieved were 0-1000 mu g/L for Cr(III) and 0-900 mu g/L for Cr(V T), with a relative standard deviation of 4.4% and 7%, respectively. T he detection limits obtained were 14.3 and 1.4 mu g/L for Cr(III) and Cr(VI) respectively. The proposed method was applied with good results to the determination of both analytes in water samples from chrome-pl ating baths.