C. Krettek et al., ACCURACY OF INTRAMEDULLARY TEMPLATES IN FEMORAL AND TIBIAL RADIOGRAPHS, Journal of bone and joint surgery. British volume, 78B(6), 1996, pp. 963-964
The accuracy of templates used for the preoperative planning of the fi
xation of intramedullary fractures depends on radiological magnificati
on. To study the accuracy of these templates, we randomly selected 100
femoral and 100 tibial radiographs taken after stabilisation by an in
tramedullary nail using a standard technique. We then compared the kno
wn nail length with the corresponding measurements on the radiographs.
The mean magnification factor for the femur was 9% and for the tibia
7%; these differ considerably from the range of magnification of the m
anufacturers' templates (femur, 15% to 17%; tibia 10% to 15%). We conc
lude that templates are unreliable for the selection of implant length
and that this should be done by intraoperative measurements.