B-1-INSENSITIVE, SINGLE-SHOT LOCALIZATION AND WATER SUPPRESSION

Citation
Ra. Degraaf et al., B-1-INSENSITIVE, SINGLE-SHOT LOCALIZATION AND WATER SUPPRESSION, Journal of magnetic resonance. Series B, 113(1), 1996, pp. 35-45
Citations number
25
Categorie Soggetti
Physics, Atomic, Molecular & Chemical
ISSN journal
10641866
Volume
113
Issue
1
Year of publication
1996
Pages
35 - 45
Database
ISI
SICI code
1064-1866(1996)113:1<35:BSLAWS>2.0.ZU;2-U
Abstract
A technique for single-shot, localized, water-suppressed in vivo H-1 N MR with high immunity to B-1 inhomogeneities is described. The single- shot localization is based on complete outer-volume suppression which is achieved with optimized, adiabatic BIR-4 pulses in combination with B-0 gradient modulation. B-1-insensitive water suppression is achieve d with BISTRO, a train of selective adiabatic full-passage pulses inte rleaved with B-0 gradient crushers. Computer simulations based on the Bloch equations, experimental images of the localized volume, and in v ivo H-1 spectroscopy experiments demonstrate the excellent spatial loc alization and water suppression. (C) 1996 Academic Press, Inc.