IN-VITRO EVALUATION OF LINEAR MOTOR-DRIVEN TOTAL ARTIFICIAL-HEART

Citation
M. Kobayashi et al., IN-VITRO EVALUATION OF LINEAR MOTOR-DRIVEN TOTAL ARTIFICIAL-HEART, Artificial organs, 20(12), 1996, pp. 1320-1324
Citations number
8
Categorie Soggetti
Engineering, Biomedical
Journal title
ISSN journal
0160564X
Volume
20
Issue
12
Year of publication
1996
Pages
1320 - 1324
Database
ISI
SICI code
0160-564X(1996)20:12<1320:IEOLMT>2.0.ZU;2-M
Abstract
This paper deals with the in vitro evaluation of a newly developed lin ear motor-driven total artificial heart (linear TAH). The linear TAHs have been developed and evaluated in mock testing and acute animal exp eriments. The new linear TAH was made based on experience with acute a nimal experiments. The maximum static thrust of the new linear pulse m otor is 146 N with exciting current of 1.69 A. The weight and the volu me of this linear TAH are 1.9 kg and 560 mi, respectively. The linear TAH has a kinetic thrust of 85 to 43 N and provides the maximum flow r ate of 6.8 L/min in mock circulatory testing.