K. Sumiyoshi et al., DETERMINATION OF CYANIDE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY USING POSTCOLUMN DERIVATIZATION WITH O-PHTHALALDEHYDE, Journal of chromatography, 690(1), 1995, pp. 77-82
A simple, highly sensitive and selective method for the determination
of cyanide ion by high-performance liquid chromatography (HPLC) with p
ostcolumn fluorescence derivatization is described. Cyanide ion was se
parated on an ion-exclusion chromatographic column and derivatized wit
h two reagents, ammonium ion in the mobile phase and o-phthalaldehyde
in the postcolumn reagent solution at pH 9.5. Sulfite ion, 2-mercaptoe
thanol and 3-mercaptopropionic acid did not interfere with the analysi
s. The average recovery of cyanide from spiked environmental water sam
ples (river water) was 99.0% (n = 6). The relative standard deviation
was 2.5% (based on peak area) for cyanide standard solutions (0.1 mu g
/ml, n = 10). The linear determination range was from 2.5 ng/ml to 1 m
u g/ml. The detection limit was 0.1 ng/ml (signal-to-noise ratio = 3,
injection volume 200 mu l).