S. Mattke et al., COMPARISON OF A UNIPOLAR DEFIBRILLATION SYSTEM WITH A DUAL LEAD SYSTEM USING AN ENLARGED DEFIBRILLATION ANODE, PACE, 19(12), 1996, pp. 2083-2088
The unipolar system for transvenous defibrillation, consisting of a si
ngle right ventricular lead as the cathode and the device shell as ano
de, has been shown to combine low defibrillation thresholds (DFTs) and
simple implantation techniques. We compared the defibrillation effica
cy of this system with the defibrillation efficacy of a dual lead syst
em with a 12-cm long defibrillation anode placed in the left subclavia
n vein. The data of 38 consecutive patients were retrospectively analy
zed. The implantation of an active can system wets attempted in 20 pat
ients (group 1), and of the dual lead system in 18 patients (group 2).
Both groups had comparable demographic data, cardiac disease, ventric
ular function, or clinical arrhythmia. The criterion for successful im
plantation was a DFT of less than or equal to 24 J. This criterion was
met in all 18 patients of group 2. The active can system could not be
inserted in 3 of the 20 group 1 patients because of a DFT > 24 J. In
these patients, the implantation of one (n = 2) or two (n = 2) additio
nal transvenous leads was necessary to achieve a DFT less than or equa
l to 24 J. The DFTs of the 17 successfully implanted group 1 patients
were not significantly different from the 18 patients in group 2 (12.3
+/- 5.7 J vs 10.8 +/- 4.8 J). The defibrillation impedance was simila
r in both groups (50.1 +/- 6.1 Omega vs 48.9 +/- 5.2 Omega). In group
1, both operation duration (66.8 +/- 17 min vs 80.8 +/- 11 min; P < 0.
05) and fluoroscopy time (3.3 +/- 2.1 min vs 5.7 +/- 2.9 min; P < 0.05
) were significantly shorter. Thus, the active can system allows relia
ble transvenous defibrillation and a marked reduction of operation dur
ation and fluoroscopy time. The dual lead system, with an increased su
rface area defibrillation anode, seems to be a promising alternative f
or active can failures.