Co. Thompson et al., MICELLAR ELECTROKINETIC CAPILLARY CHROMATOGRAPHIC DETERMINATION OF ARTIFICIAL SWEETENERS IN LOW-JOULE SOFT DRINKS AND OTHER FOODS, Journal of chromatography, 694(2), 1995, pp. 507-514
A rapid method for the determination of artificial sweeteners in low-J
oule soft drinks and other foods by micellar electrokinetic capillary
chromatography (MEKC) is described. Caffeine, benzoic acid and sorbic
acid, which are often added to soft drinks, can also be determined wit
h this procedure. The artificial sweeteners, aspartame, saccharin, ace
sulfame-K, alitame and dulcin, and the other food additives are well s
eparated in less than 12 min using an uncoated fused-silica capillary
column with a buffer consisting of 0.05 M sodium deoxycholate, 0.01 M
potassium dihydrogenorthophosphate, 0.01 M sodium berate operating at
20 kV. Dehydroacetic acid was used as the internal standard for the de
terminations. The levels of artificial sweeteners, preservatives and c
affeine were in good agreement with those determined by the high-perfo
rmance liquid chromatographic (HPLC) procedure currently used in our L
aboratory. The MEKC procedure has the same order of repeatability, is
faster and less costly to operate than the HPLC method.