VALIDATION OF THE ANALYTICAL LINEARITY AND MASS DISCRIMINATION CORRECTION MODEL EXHIBITED BY A MULTIPLE COLLECTOR INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETER BY MEANS OF A SET OF SYNTHETIC URANIUM ISOTOPE MIXTURES
Pdp. Taylor et al., VALIDATION OF THE ANALYTICAL LINEARITY AND MASS DISCRIMINATION CORRECTION MODEL EXHIBITED BY A MULTIPLE COLLECTOR INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETER BY MEANS OF A SET OF SYNTHETIC URANIUM ISOTOPE MIXTURES, Journal of analytical atomic spectrometry, 10(5), 1995, pp. 395-398
A set of commercially available synthetically prepared mixtures of ura
nium isotopes (IRMM 072) were used to assess the linearity and mass di
scrimination exhibited by a multiple collector ICP-MS instrument. Meas
urements of eight uranium isotopic reference materials with U-233:U-23
5 isotope abundance ratios ranging from 1 to 0.01 were made. Linear, p
ower law and exponential relationships between instrumental mass discr
imination and mass difference were evaluated. The power law and expone
ntial function resulted in the best correction. The corrected isotope
abundance ratios for U-233:U-235 and U-233:U-238 did not differ signif
icantly from the certified values at the 3 x 10(-4) level.