Ak. Cree et Sj. Ruff, USE OF A TALL EXTRACTOR TO REMOVE A BROKEN TAP WITHIN A GROSSE-KEMPF INTRAMEDULLARY NAIL - CASE-REPORT, The journal of trauma, injury, infection, and critical care, 38(5), 1995, pp. 802-803
A tap extractor was used to dislodge a broken tap that prevented remov
al of an intramedullary nail. This was imperative because of metal inc
ompatibility and corrosion. Correct use of extraction equipment should
allow atraumatic retrieval of intramedullary fixation. The proper tec
hnique for retapping a thread in an intramedullary nail is described.