A TABLE OF PHANTOM SCATTER FACTORS OF PHOTON BEAMS AS A FUNCTION OF THE QUALITY INDEX AND FIELD SIZE

Citation
P. Storchi et Jjm. Vangasteren, A TABLE OF PHANTOM SCATTER FACTORS OF PHOTON BEAMS AS A FUNCTION OF THE QUALITY INDEX AND FIELD SIZE, Physics in medicine and biology, 41(3), 1996, pp. 563-571
Citations number
10
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
00319155
Volume
41
Issue
3
Year of publication
1996
Pages
563 - 571
Database
ISI
SICI code
0031-9155(1996)41:3<563:ATOPSF>2.0.ZU;2-F
Abstract
Phantom scatter factors, for square fields of various sizes, have been determined at a fixed reference depth of 10 cm, separately in differe nt institutes, for a large number of linear accelerators under the aus pices of the Netherlands Commission on Radiation Dosimetry. The method used for these measurements has been described in a previous paper. T he present article describes the conversion of the measured values int o a comprehensive and consistent data set, that gives the phantom scat ter factor as a function of field size (from 4 cm up to 40 cm) and qua lity index (from 0.600 up to 0.800).