IN-VITRO MEASUREMENT OF SILICONE BLEED FROM BREAST IMPLANTS

Citation
Lt. Yu et al., IN-VITRO MEASUREMENT OF SILICONE BLEED FROM BREAST IMPLANTS, Plastic and reconstructive surgery, 97(4), 1996, pp. 756-764
Citations number
16
Categorie Soggetti
Surgery
ISSN journal
00321052
Volume
97
Issue
4
Year of publication
1996
Pages
756 - 764
Database
ISI
SICI code
0032-1052(1996)97:4<756:IMOSBF>2.0.ZU;2-0
Abstract
A method to measure gel bleed from intact silicone gel-filled breast i mplants was developed. This nondestructive technique permits accurate and reproducible serial measurements of silicone bleed from smooth wal l breast implants (n = 10) under simulated physiologic conditions in v itro. Gel bleed rates from new low bleed gel-filled implants and intac t explants (unbarriered, low bleed, double lumen) were determined. The se results demonstrate the reliability of this method to quantify sili cone gel bleed and may permit a meaningful comparison of bleed rates f rom implants in the future.