Sw. Lemire et Kl. Busch, CALIBRATION POINT FOR LIQUID SECONDARY-ION MASS-SPECTROMETRY TANDEM MASS-SPECTRA MEASURED WITH AN EBQQ HYBRID MASS-SPECTROMETER, Journal of mass spectrometry., 31(3), 1996, pp. 280-288
A protocol for establishing standard instrument conditions for the mea
surement of product ion tandem mass spectra from parent ions generated
by liquid secondary ion mass spectrometry (LSIMS) is presented. The i
on at m/z 319 (C+) from 2-butyltriphenylphosphonium bromide is selecte
d as the parent ion and then is subjected to collision-induced dissoci
ation. The relative abundances of product ions at m/z 263 and 183 are
monitored as a function of collision energy while keeping the target g
as pressure constant. The collision energy at which these two most abu
ndant product ions are equal is defined as the calibration point at th
at collision gas pressure. LSIMS tandem mass spectra acquired at the c
alibration point show good reproducibility for several model compounds
. Spectra obtained at the calibration point at two different collision
gas pressures, representing multiple and single collision conditions,
also show a high degree of similarity. Use of this protocol may aid i
n the establishment of a tandem mass spectral library.