X-RAY-IMAGING SPECTROSCOPY OF AN ELECTRON-CYCLOTRON-RESONANCE DISCHARGE

Citation
U. Lehnert et G. Zschornack, X-RAY-IMAGING SPECTROSCOPY OF AN ELECTRON-CYCLOTRON-RESONANCE DISCHARGE, Review of scientific instruments, 67(3), 1996, pp. 1264-1266
Citations number
7
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
ISSN journal
00346748
Volume
67
Issue
3
Year of publication
1996
Part
2
Pages
1264 - 1266
Database
ISI
SICI code
0034-6748(1996)67:3<1264:XSOAED>2.0.ZU;2-X
Abstract
A pinhole camera with an x-ray sensitive charge-coupled device array w as used to analyze x-rays emitted from an electron cyclotron resonance (ECR) discharge. The camera was operated at a temperature of 140 K, a llowing single photon detection at integration times up to several hou rs. This technique simultaneously allows spectroscopic investigations as well as x-ray imaging. Thus, the geometrical size and x-ray emissio n profiles from an ECR plasma were measured. (C) 1996 American Institu te of Physics.