ION TRAJECTORY CALCULATIONS FOR A QUADRUPOLE-ION-TRAP REFLECTRON-TIME-OF-FLIGHT HYBRID INSTRUMENT - EFFECTS OF THE INITIAL RF-PHASE AND THETRAPPING TIME ON AN ION BUNCH PRODUCED FROM A MOLECULAR-BEAM
U. Wilhelm et al., ION TRAJECTORY CALCULATIONS FOR A QUADRUPOLE-ION-TRAP REFLECTRON-TIME-OF-FLIGHT HYBRID INSTRUMENT - EFFECTS OF THE INITIAL RF-PHASE AND THETRAPPING TIME ON AN ION BUNCH PRODUCED FROM A MOLECULAR-BEAM, Rapid communications in mass spectrometry, 10(4), 1996, pp. 473-477
SIMION calculations are compared with experimental observations of ion
bunch parameters such as average kinetic energy, time-of-flight, and
mass resolution as a function of the storage time and the initial phas
e in the quadrupole ion trap of a trap-reflectron hybrid instrument (T
RAPTOF). It is shown that the proper choice of pr frequency and inital
phase results in favourable experimental conditions even without the
use of a buffer gas inside the trap, Synchronization of all events in
the instrument relative to the RF trapping field is the most important
prerequisite.