Ag. Casado et al., ESTIMATE OF GAS-CHROMATOGRAPHIC BLANKS - APPLICATION TO DETECTION LIMITS EVALUATION AS RECOMMENDED BY IUPAC, Journal of chromatography, 726(1-2), 1996, pp. 133-139
Citations number
24
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
A new method to obtain the signal associated with a chromatographic bl
ank is presented for inclusion within calibration procedures, Signal a
ssigned to the blank is obtained by direct integration of the backgrou
nd noise by using extrapolated values of the base-peak width at concen
trations different to the 'zero concentration'. Thus, detection limits
which are better adjusted to a statistical evaluation are implemented
. These limits are more in line with IUPAC recommendations.