RECENT RESULTS WITH A 62 G GE CRYOGENIC DARK-MATTER DETECTOR

Citation
T. Shutt et al., RECENT RESULTS WITH A 62 G GE CRYOGENIC DARK-MATTER DETECTOR, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 370(1), 1996, pp. 165-167
Citations number
7
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
370
Issue
1
Year of publication
1996
Pages
165 - 167
Database
ISI
SICI code
0168-9002(1996)370:1<165:RRWA6G>2.0.ZU;2-3
Abstract
We report on recent results with a 62 g Ge cryogenic detector that com bines simultaneous ionization and phonon measurements to allow rejecti on of radioactive background in a dark matter experiment. This detecto r has FWHM baseline resolutions of 500 eV in phonons and 1200 eV in io nization. A simplified charge collection scheme has been employed to i mprove the efficiency of the background rejection. Based on the succes s of this detector we are currently building six 165 g detectors as pa rt of the CDMS experiment.