Sw. Mackie et al., CLINICAL COMPARISON OF THE KEELER-PULSAIR-2000, AMERICAN OPTICAL MKIIAND GOLDMANN APPLANATION TONOMETERS, Ophthalmic & physiological optics, 16(2), 1996, pp. 171-177
The aim of our study was to evaluate the performance of both the Keele
r Pulsair 2000 and the American Optical (AO) MkII non-contact tonomete
rs (NCT) and compare these to the reference Goldmann standard using th
e same group of patients. Forty-five patients (89 eyes) receiving medi
cal treatment for primary open angle glaucoma had their intraocular pr
essure (IOP) measured with each instrument in a random order using fiv
e experienced observers. In the IOP range of the sample (6-27 mmHg) th
e difference between means for each tonometer was small. The Pulsair 2
000 (mean = 19.06 mmHg, SD 6.28) read slightly higher than Goldmann (m
ean = 18.01 mmHg, SD 4.88), whereas the AO MkII read slightly lower (m
ean = 16.27 mmHg, SD 5.93). However, all differences were statisticall
y significant (P < 0.001, repeat measures ANOVA), Correlation coeffici
ents for Pulsair 2000 versus Goldmann was r = 0.82 and for AO MkII ver
sus Goldmann was r = 0.85. In addition, the repeatability for each NCT
was assessed using 10 consecutive measurements on a further 10 subjec
ts. The Pulsair 2000 showed significantly greater variation of IOP (me
an SD 3.43 mmHg) than the AO MkII (mean SD 1.76 mmHg), confirming the
need for taking at least four readings per eye with the Pulsair 2000.
These variations are greater than that generally observed with Goldman
n, and possible explanations why this occurs are discussed. In conclus
ion, subject to these limitations, both NCTs should be useful for meas
uring IOP as part of a screening protocol for glaucoma.