E. Anachkova et al., CALIBRATING AND TESTING TISSUE-EQUIVALENT PROPORTIONAL-COUNTERS WITH AR-37, Radiation and environmental biophysics, 33(4), 1994, pp. 353-364
Citations number
26
Categorie Soggetti
Biophysics,"Radiology,Nuclear Medicine & Medical Imaging","Environmental Sciences
A method for testing and calibrating tissue equivalent proportional co
unters with Ar-37 is described. Ar-37 is produced by exposure of argon
in its normal isotope composition to thermal neutrons. It is shown th
at - up to volume ratios of 0.01 of argon to the tissue equivalent gas
- there is no appreciable effect of the argon admixture on the functi
on of the proportional counter. Conventional calibration methods with
characteristic x-rays or with alpha-particles require modifications of
the detectors, and they test only small sub-volumes in the counters.
In contrast, argon permits calibrations and tests of the resolution th
at are representative for the entire counter volume and that do not re
quire changes in detector construction. The method is equally applicab
le to multi-element proportional counters; it is here exemplified by i
ts application to a long cylindrical counter of simplified design that
is part of such a multi-element configuration.