M. Stancikova et al., MICROPLATE ASSAY FOR HYALURONAN USEFUL IN CLINICAL-PRACTICE, Collection of Czechoslovak Chemical Communications, 59(10), 1994, pp. 2331-2342
A microplate assay for determination of hyaluronan (HA) has been devel
oped. In principle, HA in the sample and biotinylated HA compete for b
inding to hyaluronan binding protein (HABP) coated to the plate. The a
mount of the microplate-bound biotinylated HA is quantified with avidi
n peroxidase complex. This assay is useful for 10 - 500 mug/l with < 2
% relative error with 50% inhibition at 80 mug/l. The assay is simple,
sensitive and particularly suitable for the determination of HA in se
rum and other biological fluids. The assay correlated with only commer
cially available radioisotope kit, Pharmacia HA-50, with correlation c
oefficient 0.92 at p < 0.001. Serum HA levels of patients with osteoar
thritis, rheumatoid arthritis and healthy donors determined by this as
say correspond to those previously determined by different methods. Th
ese results suggest that this assay is suitable for clinical use.