S. Dadfarnia et Cw. Mcleod, ONLINE TRACE ENRICHMENT AND DETERMINATION OF URANIUM IN WATERS BY FLOW-INJECTION INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY, Applied spectroscopy, 48(11), 1994, pp. 1331-1336
A flow injection system incorporating a microcolumn of activated alumi
na was combined with inductively coupled plasma mass spectrometry for
on-line trace enrichment and determination of uranium in surface water
s and sea water. Deposition of uranium was effected with the use of a
basic alumina microcolumn, and injection of nitric acid (250 mu L, 2M)
served to elute retained species to the ICP. A sample volume of 14 mL
resulted in a preconcentration factor of 40, and precision at the 50-
ng 1(-1) level was 4.5% (RSD). The procedure was applied to mineral wa
ters, river water, and sea water, and accuracy was assessed through ei
ther analysis of certified reference waters or recovery experiments.