Ea. Barclay et al., COMPARISON OF 2 AUTOMATED QUANTITATIVE IMMUNOASSAYS FOR THE DETERMINATION OF C-REACTIVE PROTEIN CONCENTRATIONS, Journal of Clinical Pathology, 47(12), 1994, pp. 1119-1120
Two quantitative, automated methods for the determination of C reactiv
e protein (CRP) were compared: turbidimetry (Cobas Fara II, Roche, Wel
wyn Garden City, UK) and fluorescence polarisation TDx, Abbott, Woking
ham, UK). One hundred and twenty routine serum samples submitted for m
easurement of CRP were tested using both procedures. The results were
compared using regression line analysis and showed a high degree of co
rrelation (r(2) = 0.99, X coefficient = 1.01, constant = 0.11). C reac
tive protein can be accurately measured using the automated turbidimet
ric method which can be recommended as an alternative to fluorescence
polarisation.