Pj. Vandermerwe et al., SIMULTANEOUS QUANTIFICATION OF EPHEDRINES IN URINE BY HIGHPERFORMANCELIQUID-CHROMATOGRAPHY, Journal of chromatography B. Biomedical applications, 661(2), 1994, pp. 357-361
Citations number
10
Categorie Soggetti
Chemistry Analytical
Journal title
Journal of chromatography B. Biomedical applications
A high-performance liquid chromatographic method for the simultaneous
determination of ephedrine, pseudoephedrine, phenylpropanolamine, norp
seudoephedrine and methylephedrine in urine is described. Samples were
extracted at basic pH into diethyl ether and then back extracted into
a small volume of acetic acid. These extracts were chromatographed on
a Phase Sep Spherisorb ODS 1 column with tetraethylammoniumphosphate-
methanol mobile phase and the analytes were detected by UV absorbance
at 214nm. Calibration curves were linear over the range 0.5-100 mu g/m
l for each analyte. The selectivity of the method was demonstrated for
several drugs found with the ephedrines in pharmaceutical formulation
s. This method has an excellent accuracy, precision and recovery for a
ll the ephedrine at the cut-off concentrations as set by the IOC for a
positive doping case and can thus be used to quantify these ephedrine
s when present in the urine obtained from sportspersons.