APPLICATION OF THE PHOTOIONIZATION DETECTOR FOR SINGLE DETERMINATIONSOF HYDRIDE FORMING ELEMENTS

Citation
S. Mergemeier et F. Scholz, APPLICATION OF THE PHOTOIONIZATION DETECTOR FOR SINGLE DETERMINATIONSOF HYDRIDE FORMING ELEMENTS, Fresenius' journal of analytical chemistry, 350(12), 1994, pp. 659-661
Citations number
6
Categorie Soggetti
Chemistry Analytical
ISSN journal
09370633
Volume
350
Issue
12
Year of publication
1994
Pages
659 - 661
Database
ISI
SICI code
0937-0633(1994)350:12<659:AOTPDF>2.0.ZU;2-M
Abstract
The application of a photoionization detector in conjunction with a fl ow-through system is described for the determination of arsenic, antim ony, selenium, tellurium, bismuth, tin, sulfide and ammonia. The analy tical procedure is based upon a preliminary generation of the hydride compounds, their subsequent transfer to the gas phase and finally thei r detection in the photoionization detector. Linear calibration curves were obtained over three to four orders of magnitude with the followi ng 3s-detection limits: 4 ng/ml As, 0.3 ng/ml Sb, 0.5 ng/ml Se, 3.0 ng /ml Te, 0.1 mug/ml Bi, 0.5 mug/ml Sn, 10.0 ng/ml S2-, 0.5 mug/ml NH3.