Mo. Leach et al., QUALITY ASSESSMENT IN IN-VIVO NMR-SPECTROSCOPY .3. CLINICAL-TEST OBJECTS - DESIGN, CONSTRUCTION, AND SOLUTIONS, Magnetic resonance imaging, 13(1), 1995, pp. 131-137
Based on the requirements of test protocols developed to evaluate clin
ical MRS single slice and volume localisation sequences, two clinical
test objects, STO1 and STO2 have been developed. The properties of a r
ange of potential construction materials have been assessed, demonstra
ting that the water/Perspex interface produced minimum susceptibility
effects. The design of the objects has been evaluated in trials on dif
ferent magnetic resonance instruments, with size and loading being adj
usted to allow use on currently available equipment. Appropriate test
solutions for P-31 and H-1 measurements have been developed and their
properties evaluated.