CRYOSTAT FOR NEUTRON INVESTIGATIONS ON A LARGE-VOLUME OF LIQUID-HE-4 BETWEEN 5.2-K AND 1.2-K AT PRESSURES UP TO 30-BARS

Citation
Iv. Bogoyavlenskii et al., CRYOSTAT FOR NEUTRON INVESTIGATIONS ON A LARGE-VOLUME OF LIQUID-HE-4 BETWEEN 5.2-K AND 1.2-K AT PRESSURES UP TO 30-BARS, Cryogenics, 34, 1994, pp. 255-257
Citations number
4
Categorie Soggetti
Physics, Applied",Thermodynamics
Journal title
ISSN journal
00112275
Volume
34
Year of publication
1994
Supplement
ICEC
Pages
255 - 257
Database
ISI
SICI code
0011-2275(1994)34:<255:CFNIOA>2.0.ZU;2-M
Abstract
Investigation of liquid helium-4 by neutrons scattering methods demand s special cryogenic equipment. The experimental conditions require a p rolonged exhibition of the large liquid helium sample. Besides, helium -4 is one of the difficult subjects for cooling, as it possesses high heat content, the thermal capacity peak in lambda-point, and the therm al conductivity in the superfluid state. A cryostat for the inelastic neutron scattering research is described. Cryostat permit to operate w ith the large liquid helium sample (2600cm(3)) at the temperature rang e 5.2 - 1.2K and pressure up to 30 bars.