CALCULATION OF SENSITIVITY CORRECTION FACTORS FOR SURFACE COIL MRS

Citation
Vl. Doyle et al., CALCULATION OF SENSITIVITY CORRECTION FACTORS FOR SURFACE COIL MRS, Magnetic resonance in medicine, 33(1), 1995, pp. 108-112
Citations number
14
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
07403194
Volume
33
Issue
1
Year of publication
1995
Pages
108 - 112
Database
ISI
SICI code
0740-3194(1995)33:1<108:COSCFF>2.0.ZU;2-D
Abstract
Quantification of MRS signals obtained with surface coils is difficult due to the inhomogeneous response of these coils, This inhomogeneity results in the measured signal from a defined volume of interest (VoI) being spatially dependent. To account for the sensitivity variation w ith position from the surface coil, we have developed a method of calc ulating correction factors for defined VOIs based on an experimentally obtained 3D sensitivity coil map. These factors may then be applied t o spectra obtained from these VOIs to accurately take into considerati on the varying coil sensitivity resulting in a reduction of measured s ignal. This method is demonstrated here to be able to correct for the inhomogeneity of surface coils over a range of two coil radii to withi n 4% accuracy.