A. Welch et al., A COMPARISON OF GD TC VERSUS TC/TL SIMULTANEOUS TRANSMISSION AND EMISSION IMAGING USING BOTH SINGLE AND TRIPLE DETECTOR FAN-BEAM SPECT SYSTEMS/, IEEE transactions on nuclear science, 41(6), 1994, pp. 2779-2786
In this study, we compare the accuracy of the transmission and emissio
n maps, produced using both a Gd/Tc and a Tc/Tl transmission-emission
source combination. In both cases, the transmission and emission data
are acquired simultaneously. This results in contamination of the data
acquired in the lower energy window by events from the higher energy
isotope. We use standard methods to correct for this contamination whe
n using a Gd/Tc combination and propose a new method for use with the
Tc/Tl combination. The Tc/Tl combination yields more accurate transmis
sion images than the Gd/Tc combination. However, when a single factor
is used to scale the attenuation coefficients measured using the trans
mission source to those appropriate for the emission isotope/energy, t
he resulting error in the attenuation map used for attenuation correct
ion is larger for the Tc/Tl combination than for the Gd/Tc combination
. For this study, neither combination yielded emission maps which were
significantly superior (in terms of reconstructed emission ratios) to
the other.